Laboratory Testing of Hydrocephalus Shunts–Conclusion of ... | 合法醫療器材資訊網
16 models of valves, currently in use in the U.K., have been tested long-term in the U.K. Shunt Evaluation Laboratory according to the protocol based on the new ISO 7197 standard. Valves tested were:
Medtronic PS Medical: Delta Valve, Flow Control and Lumbo-Peritoneal Shunt
Heyer-Schulte Nero-Care: In-line, Low Profile and Pudenz Flushing Valve
Codman: Codman-Hakim Programmable, Hakim Precision, Accu-Flo, Holter, Uni-Shunt, and siphon-preventing device – SiphonGuard
NMT: Orbis-Sigma Valve, Omni-Shunt and Hakim Valve
Sophysa: Sophy Programmable Valve
Radionics: Contour-Flex Valve.
The majority of the valves had a non-physiologically low hydrodynamic resistance (with the exception of Orbis-Sigma, PS Lumbo-Peritoneal and Heyer-Schulte In-Line). This may result in overdrainage both related to posture and during noc...