16 models of valves, currently in use in the U.K., have been tested long-term in the U.K. Shunt Evaluation Laboratory according to the protocol based on the new ISO 7197 standard. Valves tested were: Medtronic PS Medical: Delta Valve, Flow Control and Lumbo-Peritoneal Shunt, Heyer-Schulte Nero-Care: In-line, Low Profile and Pudenz Flushing Valve, Codman: Codman-Hakim Programmable, Hakim Precision, Accu-Flo, Holter, Uni-Shunt, and siphon-preventing device -- SiphonGuard, NMT: Orbis-Sigma Valve, Omni-Shunt and Hakim Valve, Sophysa: Sophy Programmable Valve, Radionics: Contour-Flex Valve. The majority of the valves had a non-physiologically low hydrodynamic resistance (with the exception of Orbis-Sigma, PS Lumbo-Peritoneal and Heyer-Schulte In-Line). This may result in overdrainage both related to posture and during nocturnal cerebral vasogenic waves. A long distal catheter increases the resistance of these valves by 1...