Recalls and alerts more than 4 years old are automatically archived. While this information can still be accessed in the database, it has not been altered or updated since it was archived. Web pages that are archived on the Web are not subject to the Government of Canada Web Standards. As per the Communications Policy of the Government of Canada[1], you can request alternate formats by contacting us[2].
ReasonSiemens Healthcare Diagnostics has observed that specific lots of the Siemens APTT Reagent, Dade Actin FSL Activated PTT Reagent, show an increase in heparin sensitivity over the shelf life. This increased heparin sensitivity prolongs the activated partial thromboplastin times (APTT) which may lead to a decision to decrease the anticoagulation dosage, increasing the risk of thromboembolism.
Affected products Dade Actin FSL Activated PTT Reagent (2013-02-25) ...Class 2 Device Recall Dade(R) Actin(R) FSL Activated PTT ... | 合法醫療器材資訊網
MAUDE Adverse Event Report | 合法醫療器材資訊網
奇美醫院病理中心檢驗資訊表 | 合法醫療器材資訊網
Dade Actin FSL Activated PTT Reagent (February 25 | 合法醫療器材資訊網
Dade® Actin® FSL Activated PTT Reagent | 合法醫療器材資訊網
SEED Coagulation | 合法醫療器材資訊網
Dade® Actin® FSL Activated PTT Reagent B4219 | 合法醫療器材資訊網
【戴德艾庭FSL活化部分凝血活每試劑】許可證字號:衛署醫器輸字第011713號