An improved microalbumin method (microALB | 合法醫療器材資訊網
由PDatta著作·2009·被引用22次—Quantitativedeterminationofalbumin(ALB)inhumanurineisimportanttoassess...AllADVIAChemistrySystemsusethesamemicroalbuminreagentpacks, ...
Quantitative determination of albumin (ALB) in human urine is important to assess kidney functions in a variety of diseases. Recently, Siemens released an improved Microalbumin assay (microALB_2) to measure urinary ALB on the automated, random access ADVIA 1650/1800, ADVIA 2400, and ADVIA 1200 Chemistry Systems. We evaluated analytical performances of this new method. All ADVIA Chemistry Systems use the same microalbumin reagent packs, microALB_2 calibrators, and commercial controls. The within-run and total CVs of the improved method with two-level BioRad Liquichek Urine Chemistry controls (approximately 2 and 9 mg/dl ALB) and a urine pool (approximately 29 mg/dl ALB) on all ADVIA Chemistry systems were <4.1 and <6.1%, respectively (40 replicates per sample). The analytical range/linearity of the method (all ADVIA systems) was from 0.3 mg/dl to theALB concentration in the highest level of calibrator (approxim...