由SSKhan著作·1990·被引用27次—DifferencesinHancockandCarpentier-Edwardsporcinexenograftaorticvalvehemodynamics.Effectofvalvesize.Circulation.1990Nov;82(5Suppl):IV117-24.
We prospectively compared the hemodynamic performance of Hancock and Carpentier-Edwards bioprosthetic aortic valves in a randomized study of 100 patients. A total of 47 patients received the Hancock valve, and 53 received the Carpentier-Edwards valve. Mean pressure gradients were measured using micromanometer catheters and cardiac outputs by thermodilution. Multiple measurements were made in each patient with atrial pacing, volume infusion, and inotropic drugs for a total of 319 observations. The pressure gradients and Gorlin valve areas showed significant scatter caused by both flow-dependent and patient-dependent variability. Mean transvalvular pressure gradients were therefore compared after adjustment for flow rate and for random interpatient differences using analysis of variance and covariance. Pressure gradients were lower and Gorlin valve areas larger for the Hancock valve than for the Carpentier-Edwards val...