MAUDE Adverse Event Report | 合法醫療器材資訊網
Concomitant products: cook memory ii double lumen extraction basket, balloon catheter (unknown type), forceps (unknown type), boston scientific hydra jagwire 0.035 inch.Investigation evaluation: our evaluation of the product said to be involved confirmed that the metal side ramp in the distal end of the catheter is not in the lumen.The metal side ramp was returned in a separate clear bag.The outer diameter of the metal side ramp measured to be within specification.During a visual inspection, it was noted that a dried green/yellow substance was on the inside and outside of the howell biliary introducer catheter.This substance was removed through the decontamination process.A small kink is noted on the skived portion of the distal catheter lumen.There is also a yellow mark approximately 1.5 cm in length just ...
Cook | 合法醫療器材資訊網
G22920 | 合法醫療器材資訊網
Howell Biliary Introducer Needle | 合法醫療器材資訊網
MAUDE Adverse Event Report | 合法醫療器材資訊網
Products | 合法醫療器材資訊網
“曲克”郝衛爾膽道導入器和針“COOK” HOWELL BILIARY ... | 合法醫療器材資訊網
“曲克”郝衛爾膽道導入器和針“COOK” HOWELL BILIARY ... | 合法醫療器材資訊網
【“曲克”郝衛爾膽道導入器和針】許可證字號:衛署醫器輸字第021874號
器材品名:“曲克”郝衛爾膽道導入器和針許可證字號:衛署醫器輸字第021874號註銷狀態:註銷日期:註銷理由:有效日期:2020/12/23發...