Thelmmbflexreagentcartridgeinstructionsforuselimitationsofproceduresection...DIMENSION®LMMBMASSCREATINEKINASEMBISOENZYMEFLEX® ...
Analysis of the instrument and instrument data indicates that the cause for the discordant elevated lmmb result is unknown.However, the siemens headquarters support center evaluated the information made available.The determination was that the elevated mmb value is patient specific, repeatable, and discordant with the cki value.Testing on an alternate instrument system recovers a value within the normal range, congruent with the cki value.This data is consistent with non-specific binding interference.The lmmb flex reagent cartridge instructions for use limitations of procedure section states: "patient samples may contain heterophilic antibodies that could react in immunoassays to give falsely elevated or depressed results.This assay has been designed to minimize interference from heterophilic antibodi...