The suctioning efficiency and trauma-producing characteristics of five commercially available tracheobronchial suction catheters (Pharmaseal Tri-Flo, NCC Gentle-Flo, Argyle Aero-Flo, Argyle Dual Side-Hole, and Pharmaseal Whistle-Tip) were experimentally evaluated in anesthetized healthy dogs. The tendency of catheters to invaginate or "grab" tracheobronchial mucosa was observed with a bronchofiberscope during suctioning. Mucosal grabbing was seldom seen even at high (greater than 300 torr) vacuum levels with the cateter tip in the trachea. All catheters were observed to invaginate mucosa in lobar and segmental bronchi, with the frequency of grabbing being a function of airway anatomy, airway size, catheter orientation, tip design, and vacuum level. Catheters with multiple side-holes appeared to invaginate mucosa less frequently than the single side-hole catheter. Repeated suctioning of anesthetized healthy dogs foll...