[Reliability of a positron emission tomography system (CTI ... | 合法醫療器材資訊網
由SWatanuki著作·2002·被引用2次—ThereliabilityofthePETsystemstronglydependsontheMTBFofDBandBC.Theimprovementinqualityofthesepartsandoptimizationoftheenvironmentin ...
Purpose: The maintenance data of a PET system (PT931/04-12 CTI Inc.) was analyzed to evaluate its reliability.
Method: We examined whether the initial performance for the system resolution and efficiency is kept. The reliability of the PET system was evaluated from the value of MTTF (mean time to failure) and MTBF (mean time between failures) for each part of the system obtained from the maintenance data for 13 years.
Results: The initial performance was kept for the resolution, but the efficiency decreased to 72% of the initial value. The 83% of the troubles of the system was for detector block (DB) and DB control module (BC). The MTTF of DB and BC were 2,733 and 3,314 days, and the MTBF of DB and BC per detector ring were 38 and 114 days. The MTBF of the system was 23 days. We found seasonal dependence for the number of troubles of DB and BC. This means that the trouble may be rel...
Siemens to Acquire CTI | 合法醫療器材資訊網
[Reliability of a positron emission tomography system (CTI ... | 合法醫療器材資訊網
正子造影技術 | 合法醫療器材資訊網
CTI sets sights on molecular imaging; debuting lower | 合法醫療器材資訊網
Multislice CT moves into PET | 合法醫療器材資訊網
The Development of a Compact Positron ... | 合法醫療器材資訊網
History and future technical innovation in positron emission ... | 合法醫療器材資訊網
Positron emission tomography: An overview | 合法醫療器材資訊網
The Development of a Compact Positron Tomograph for ... | 合法醫療器材資訊網
【正子閃爍斷層攝影系統】許可證字號:衛署醫器輸字第005890號
器材品名:正子閃爍斷層攝影系統許可證字號:衛署醫器輸字第005890號註銷狀態:已註銷註銷日期:2013/01/17註銷理由:許可證已逾...