Background: The absence of the synergic effect between beta-lactam and aminoglucoside versus Enterococcus is predicted by the presence of a high level of resistance to aminoglucoside. There are several commercial systems which include wells for the performance of this determination for streptomycin and gentamicin.
Methods: A comparative study of two commercial systems, Vitek GPS-TA and ATB STREP (Biomérieux) and disk diffusion versus the Mueller-Hinton agar dilution method used as a method of reference. Two hundred ninety strains of Enterococcus spp. from blood, wounds, pus, fluids from paracentesis and urine were studied.
Results: By the reference method, 17 presented high level of resistance only to gentamicin, 45 only to streptomycin, 46 to both and 182 were sensitive to the two antibiotics. The sensitivity of ATB STREP, Vitek GPS-TA and disk diffusion was respectively for high level...