Background and Aims: In clinical practice dialysate conductivity is typically set to 14 mS/cm. The aim of this study was to correlate the conductivity displayed on a dialysis machine panel and the dialysate sodium concentration [Na], and also to evaluate the effect of dialysate [Na] on patients interdialytic fluid gain. Patient and Methods: A total of 339 patients dialyzed with the Althin system (Drake Willock System 1000 Single Patient Delivery System, Althin Medical, mc, Portland, Oregon; N=56 machines), or the Toray system (Toray TR-321 Series Single-Patient Dialysis System, Toray Medical Co., Ltd. Tokyo, Japan; N=20 machines) were enrolled in this study. The dialysate [Na] was analyzed with Nova Electrolyte/Chemistry Analyzers (Nova CRT-13, Nova Biochemical, MA, US). Maintenance of machines was performed regularly according to the manufacturers guidelines. Results: The conduc...
Airiti Library華藝線上圖書館 | 合法醫療器材資訊網
K954987.pdf | 合法醫療器材資訊網
K964922.pdf | 合法醫療器材資訊網
Summary | 合法醫療器材資訊網
單一病人透析系統"ALTHIN" SINGLE PATIENT DIALYSATE ... | 合法醫療器材資訊網
單一病人透析系統"ALTHIN" SINGLE PATIENT DIALYSATE ... | 合法醫療器材資訊網
單一病人透析系統"CD" SINGLE PATIENT DIALYSATE ... | 合法醫療器材資訊網
單一病人透析系統"CD" SINGLE PATIENT DIALYSATE DELIVERY ... | 合法醫療器材資訊網
【單一病人透析系統】許可證字號:衛署醫器輸字第003527號