[143988968]Thecustomercontactedasiemenscustomercarecenter(ccc)andreportedthatadiscordant,falselyelevatedalbuminresultwasobtainedona ...
[143988968] The customer contacted a siemens customer care center (ccc) and reported that a discordant, falsely elevated albumin result was obtained on a patient sample on the bn prospec system. Siemens analyzed the submitted data file, system logs and the instrument service history and determined there was no evidence of instrument malfunction. Siemens is investigating the issue. Patient Sequence No: 1, Text Type: N, H10
[143988969] A discordant, falsely elevated albumin/cerebrospinal fluid (csf) result was obtained using n antiserum to human albumin lot 153936e on the bn prospec system. The albumin was used for calculating csf/albumin serum ratio. The initial falsely elevated result was not reported to the physic...