由KStoffel著作·2017·被引用47次—BiomechanicalEvaluationoftheFemoralNeckSysteminUnstablePauwelsIIIFemoralNeckFractures:AComparisonwiththeDynamicHipScrewand ...
Objectives: To evaluate the biomechanical performance of femoral neck system (FNS) in comparison with established methods for fixation of femoral neck fractures in a cadaveric model.
Methods: Twenty pairs of fresh-frozen human cadaveric femora were instrumented either with dynamic hip screw and antirotation screw (DHS-screw), DHS-blade, 3 cannulated screws (3CS) or with FNS in a partially paired design. The specimens were randomized to 2 paired treatment groups based on the bone mineral density (BMD), namely DHS-screw/DHS-blade and FNS/3CS. A reduced unstable femoral neck fracture with postero-caudal comminution, OTA/AO 31-B2.3, 70 degrees Pauwels III, was simulated by cutting 30 degrees distal and 15 degrees posterior wedges. Cyclic axial loading was applied in 16 degrees adduction, starting at 500 N and with progressive peak force increase of 0.1 N/cycle until construct failure. Axial stiffness wa...
Biomechanical Evaluation of the Femoral Neck ... | 合法醫療器材資訊網
Biomechanical Evaluation of the Femoral Neck System in ... | 合法醫療器材資訊網
Biomechanical evaluation of femoral neck fracture fixation with ... | 合法醫療器材資訊網
[PDF] Biomechanical Evaluation of the Femoral Neck System ... | 合法醫療器材資訊網
Biomechanical Evaluation of the Femoral Neck ... | 合法醫療器材資訊網
Biomechanical evaluation of locked plating fixation for ... | 合法醫療器材資訊網
Biomechanical Evaluation of Internal Fixation of Pauwels Type ... | 合法醫療器材資訊網
【“信迪思”鎖定加壓髖部骨板系統】許可證字號:衛署醫器輸字第019261號