2018年12月28日—Replacedbatterynochange.Leftbatteryoutfor1weekthenturnedpumpontogettheerrorcodeanditwasworking.Continuedtowork(plugged ...
The battery that can cause this is the RTC battery (real time clock). This is a seperate battery from the main battery. The RTC battery is located on the display board. According to the service manual, automatic test mode #1, the back up battery voltage should be BB 574-737 when plugged into outlet power. While in test mode #1, check main battery (as its values will be on the top line) MB should be 688-714. Oncethe bach up battery has been replaced/removed, the pumps configuration willl need to be changed in accordance to your institutions policy.
Also, service manual says the next possiblity is U506. If that does not fix the alarm, then to replace the CPU board.