Comparison of activated clotting times measured ... | 合法醫療器材資訊網
由JMLee著作·2018·被引用9次—Theactivatedclottingtime(ACT)isdefinedasthedurationoftimeuntilclot...TheHemochronJr.Signaturemeasurestheelapsedtimebetweenthestart...TheMedtronicACTPlus(Medtronic,Minneapolis,MN,USA)use
Among the ACT values of <400 s following heparin injection (10 samples), seven patients had both an HACT and MACT of <400 s, one patient had an HACT of <400 s and an MACT of >400 s, and two patients had an HACT of >400 s and MACT of <400 s. The bias between the two ACT devices was 14.3 s, suggesting that the HACT was higher than the MACT.
Considering the difference in the baseline ACT observed between the two devices, we also compared the following two ratios: HACT after heparinisation or protamine injection/HACT at baseline (rHACT) versus MACT after heparinisation or protamine injection/MACT at baseline (rMACT). The rHACT and rMACT were correlated (r = 0.945, p < 0.0001) () with a small bias of −0.08 (limit of agreement, −0.97 to 0.81) ().
Multiple regression analysis showed that the blood glucose concentration significantly influenced the differences between the two ACT devices...
(PDF) Comparison of activated clotting times measured using ... | 合法醫療器材資訊網
Comparison of activated clotting times measured ... | 合法醫療器材資訊網
Comparison of activated clotting times measured ... | 合法醫療器材資訊網
Comparison of activated clotting times measured using the ... | 合法醫療器材資訊網
Correlation between activated clotting time monitoring and ... | 合法醫療器材資訊網
Evaluations of Activated Clotting Time Technologies Require ... | 合法醫療器材資訊網
HEMOCHRON Activated Clotting Time Low Range ... | 合法醫療器材資訊網
Hemochron Signature ACT+ Procedure in Cardiac OR | 合法醫療器材資訊網
Point | 合法醫療器材資訊網
【漢莫克全血活化凝血時間試驗】許可證字號:衛部醫器輸字第025263號
器材品名:漢莫克全血活化凝血時間試驗許可證字號:衛部醫器輸字第025263號註銷狀態:註銷日期:註銷理由:有效日期:2023/08/13發...