Breathing through an endotracheal tube, connector, and ventilator demand valve imposes an added load on the respiratory muscles. As respiratory muscle fatigue is thought to be a frequent cause of ventilator dependence, we sought to examine the efficacy of five different ventilators in reducing this imposed work through the application of pressure support ventilation. Using a model of spontaneous breathing, we examined the apparatus work imposed by the Servo 900-C, Puritan Bennett 7200a, Engstrom Erica, Drager EV-A or Hamilton Veolar ventilators, a size 7.0 and 8.0 mm endotracheal tube, and inspiratory flow rates of 40 and 60 l/min. Pressure support of 0, 5, 10, 15, 20 and 30 cm H2O was tested at each experimental condition. Apparatus work was greater with increased inspiratory flow rate and decreased endotracheal tube size, and was lowest for the Servo 900-C and Puritan Bennett 7200a ventilators. Apparatus work fell...