We compared the annotation of the AHA and MIT databases beat-to-beat with the classification preformed by the microprocessor of a 24-hour ambulatory electrocardiographic device, based on real-time analysis. Sensitivity and positive predictive accuracy for QRS detection were 99.9% (99.9%) and 99.9% (99.8%) for the AHA database (MIT database respectively). Sensitivity and positive predictive accuracy were 99.1% (96.6%) and 98.3% (94.9%) for ventricular ectopic beats, 98.3% (91.8%) and 96.0% (63.0%) for couplets and 96.4% (74.2%) and 99.2% (41.1%) for salvoes. On 90% of the AHA tapes (MIT tapes) sensitivity and positive predictive accuracy were at least 93.8% (76.6%) and 92.7% (65.5%) for ventricular ectopic beats, at least 98.0% (96.3%) and 54.5% (0%) for couplets and at least 100% (66.6%) and 100% (0%) for salvoes. A sensitivity of 100% was achieved for ventricular ectopic beats on 56% (45%), for couplets on 90% (82%...