Biomechanical comparison of a knotless suture anchor with ... | 合法醫療器材資訊網
Purpose: To compare the biomechanical strength of knotless suture anchors and standard suture anchors in the repair of type II SLAP tears.
Methods: Five pairs of cadaveric shoulders (10 shoulders) were dissected free of soft tissue except for the glenoid labrum and long head of the biceps tendon. Type II SLAP tears were created and repaired with 1 of 2 anchors: the Mitek Lupine suture anchor or the Mitek Bioknotless suture anchor (DePuy Mitek, Raynham, MA). All specimens were preloaded to 10 N, and loaded for 25 cycles in 10 N increments to a maximum of 200 N. If specimens were still intact after 200 N, they were loaded to ultimate failure. The load at which 2 mm of gapping occurred, load to ultimate failure, mode of failure, and the number of cycles to failure were compared using the Wilcoxon signed-rank test.
Results: Load to 2-mm gapping was lower (P = .042) for knotless anchors (70 ...
6775 bioknotless SS HiRes.fnl | 合法醫療器材資訊網
5957r3 Mitek Bioknotless | 合法醫療器材資訊網
The Easiest Knot You'll Never Tie | 合法醫療器材資訊網
7150 BIOKNOTLESS | 合法醫療器材資訊網
BIOKNOTLESS™ PLUS | 合法醫療器材資訊網
Arthroscopic Bankart repair using Knotless or BioKnotless ... | 合法醫療器材資訊網
Biomechanical comparison of a knotless suture anchor with ... | 合法醫療器材資訊網
&benefits features | 合法醫療器材資訊網
DePuy Mitek Spanning Technique | 合法醫療器材資訊網
【"邁特" 可吸收免打結固定錨】許可證字號:衛署醫器輸字第009933號
器材品名:"邁特"可吸收免打結固定錨許可證字號:衛署醫器輸字第009933號註銷狀態:已註銷註銷日期:2019/11/04註銷理由:許可證...